200keV saturated lens of a TEM - Field accuracy |
Page 4 of 4
Field accuracy
The plot shows in blue the axial flux density (maximum 3.30 T) and in red the error estimate (the difference between the computations a mesh with 80000 points and with 320000 points at the mesh points of the older mesh – see papers from 2002-2004). The red curve is magnified 10000 times and the error is between +0.28 mT and –0.25 mT.
|